CYCLIC FATIGUE RESISTANCE OF VARIOUS NICKEL-TITANIUM ROTARY ENDODONTIC INSTRUMENTS AT BODY TEMPERATURE: AN IN VITRO STUDY

  • DARA H. SAEED B.D.S., M.Sc., Ph.D., Assistant Professor at the College of Dentistry - Hawler Medical University, Erbil 44001, Kurdistan Region-Iraq
  • FADI A. RAFEA B.D.S., At the College of Dentistry - Hawler Medical University, Erbil 44001, Kurdistan Region-Iraq
Keywords: Cyclic Fatigue:, Files:, Nickel-Titanium:, Body Temperature:

Abstract

Introduction: The purpose of this study was to compare the cyclic fatigue of different types of nickel-titanium rotary files at body temperature.

Methods: Four types of rotary instruments with tip size 0.25 were used: HyFlex EDM (OneFile, variable taper, Colten/Whaledent), ProTaper Gold (F2, 0.08 taper, Dentsply), XP-endo Finisher (zero tapers, FKG Dentaire) and 2Shape (TS2, 0.06 taper, Micro-Mega). Ten files for each instrument were tested for cyclic fatigue at body temperature and rotated until fracture occurred in a simulated canal with an angle curvature of 60°, a radius curvature of 5 mm, and a canal width of 1.5 mm. Four groups were formed for all instruments (total number = 40). The number of cycles to fracture and the length of fractured fragment were recorded, and statistical analysis was completed using analysis of variance with significance at a (p-value < 0.05).

Results: The cyclic fatigue resistance of the HyFlex EDM OneFile was significantly greater than the other instruments, while that of the 2Shape TS2 was significantly lower (p < 0.05).

Conclusion: Within the limitations of this study, the HyFlex EDM OneFile was more resistant to cyclic fatigue than the XP-endo Finisher, ProTaper Gold F2 and 2Shape TS2 files

Downloads

Download data is not yet available.

References

1. Arias A, Perez-Higueras JJ, de la Macorra JC. Differences in cyclic fatigue resistance at apical and coronal levels of Reciproc and WaveOne new files. J Endod. 2012;38:1244–1248.
2. Perez-Higueras JJ, Arias A, de la Macorra JC. Cyclic fatigue resistance of K3, K3XF, and twisted file nickel-titanium files under continuous rotation or reciprocating motion. J Endod. 2013;39:1585–1588.
3. McGuigan MB, Louca C, Duncan HF. Clinical decision-making after endodontic instrument fracture. Br Dent J. 2013;214:395–400.
4. Sattapan B, Nervo GJ, Palamara JE, Messer HH. Defects in rotary nickel-titanium files after clinical use. J Endod. 2000;26:161–165.
5. Pedulla E, Savio FL, Boninelli S, et al. Torsional and cyclic fatigue resistance of a new nickel-titanium instrument manufactured by electrical discharge machining. J Endod. 2016;42:156–159.
6. Bouska J, Justman B, Williamson A, et al. Resistance to cyclic fatigue failure of a new endodontic rotary file. J Endod. 2012;38:667–669.
7. Setzer FC, Bohme CP. Influence of combined cyclic fatigue and torsional stress on the fracture point of nickel-titanium rotary instruments. J Endod. 2013;39:133–137.
8. Kim HC, Kwak SW, Cheung GS, Ko DH, Chung SM, Lee W. Cyclic fatigue and torsional resistance of two new nickel-titanium instruments used in reciprocation motion: Reciproc versus WaveOne. J Endod. 2012;38:541–544.
9. Peters O, Gluskin A, Weiss R, Han J. An in vitro assessment of the physical properties of novel Hyflex nickel–titanium rotary instruments. Int Endod J. 2012;45:1027–1034.
10. Pirani C, Iacono F, Generali L, et al. HyFlex EDM: superficial features, metallurgical analysis and fatigue resistance of innovative electro discharge machined NiTi rotary instruments. Int Endod J. 2016;49:483–493.
11. Hieawy A, Haapasalo M, Zhou H, et al. Phase transformation behavior and resistance to bending and cyclic fatigue of ProTaper Gold and ProTaper Universal instruments. J Endod. 2015;41:1134–1138.
12. Uygun AD, Kol E, Topcu MK, et al. Variations in cyclic fatigue resistance among ProTaper Gold, ProTaper Next and ProTaper Universal instruments at different levels. Int Endod J. 2016;49:494–499.
13. FKG Dentaire. The XP-endo finisher brochure. Available at: http://www.fkg.ch/sites/default/files/201610_fkg_xp_endo_finisher_brochure_v2_en_web.pdf. 2016 (last accessed 10/2018).
14. MicroMega. The 2Shape brochure. Available at: http://www.denta.be/pdf/MicroMega-2shape-ENG.pdf. 2017 (last accessed 10/2018).
15. Elnaghy A. Cyclic fatigue resistance of ProTaper next nickel titanium rotary files. Int Endod J. 2014;47:1034–1039.
16. Plotino G, Grande NM, Cotti E, et al. Blue treatment enhances cyclic fatigue resistance of vortex nickel-titanium rotary files. J Endod. 2014;40:1451–1453.
17. de Hemptinne F, Slaus G, Vandendael M, et al. In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. J Endod. 2015;41:1112–1115.
18. Jamleh A, Yahata Y, Ebihara A, et al. Performance of NiTi endodontic instrument under different temperatures. Odontology. 2016;104:324–328.
19. de Vasconcelos RA, Murphy S, Carvalho CA, et al. Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. J Endod. 2016;42:782–787.
20. Dosanjh A, Paurazas S, Askar M. The Effect of temperature on cyclic fatigue of nickel-titanium rotary endodontic instruments. J Endod. 2017;43:823–826.
21. Pruett JP, Clement DJ, Carnes DL Jr. Cyclic fatigue testing of nickel-titanium endodontic instruments. J Endod. 1997;23:77–85.
22. Gündoğar M, Özyürek T. Cyclic fatigue resistance of OneShape, HyFlex EDM, WaveOne gold, and Reciproc blue nickel-titanium instruments. J Endod. 2017;43:1192–1196.
23. Larsen CM, Watanabe I, Glickman GN, He J. Cyclic fatigue analysis of a new generation of nickel titanium rotary instruments. J Endod. 2009;35:401–403.
24. Grande NM, Plotino G, Silla E, et al. Environmental temperature drastically affects flexural fatigue resistance of nickel-titanium rotary files. J Endod. 2017;43:1157–1160.
25. Plotino G, Grande NM, Cordaro M, Testarelli L, Gambarini G. A review of cyclic fatigue testing of nickel-titanium rotary instruments. J Endod. 2009;35:1469–1476.
26. Kaval ME, Capar ID, Ertas H. Evaluation of the cyclic fatigue and torsional resistance of novel nickel-titanium rotary files with various alloy properties. J Endod. 2016;42:1840–1843.
27. Özyürek T, Gündoğar M, Uslu G, et al. Cyclic fatigue resistances of Hyflex EDM, WaveOne gold, Reciproc blue and 2shape NiTi rotary files in different artificial canals. Odontology. 2018;106:408–413.
28. Cheung GS, Zhang EW, Zheng YF. A numerical method for predicting the bending fatigue life of NiTi and stainless steel root canal instruments. Int Endod J. 2011;44:357–361.
29. Keskin C, Inan U, Guler DH, Kalyoncuoğlu E. Cyclic Fatigue Resistance of XP-Endo Shaper, K3XF, and ProTaper Gold Nickel-titanium Instruments. J Endod. 2018;44:1164–1167.
30. Vaz-Garcia ES, Vieira VT, Natasha Pereira Da Silva Ferreira Petitet, et al. Mechanical Properties of Anatomic Finishing Files: XP-Endo Finisher and XP-Clean. Brazilian Den J. 2018;29:208–213.
Published
2020-01-26
How to Cite
SAEED, D. H., & RAFEA, F. A. (2020). CYCLIC FATIGUE RESISTANCE OF VARIOUS NICKEL-TITANIUM ROTARY ENDODONTIC INSTRUMENTS AT BODY TEMPERATURE: AN IN VITRO STUDY. Journal of Duhok University, 22(1), 61-66. https://doi.org/10.26682/sjuod.2020.22.1.8